A Hoffman brain phantom lesion study with the transformable HOTPET camera

  • Hossain Baghaei
  • , Hongdi Li
  • , Yuxuan Zhang
  • , Rocio A. Ramirez
  • , Shitao Liu
  • , Chao Wang
  • , Shaohui An
  • , Wai Hoi Wong

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Abstract

The High-resolution Oncologic Transformable PET (HOTPET) camera has intrinsic spatial resolution of better than 3 mm and can operate in different modes including the whole-body mode and brain mode. In the whole-body mode, HOTPET has a transverse field-of-view (FOV) of 60 cm and an axial FOV of 13 cm. In the brain mode it has a transverse FOV of 39 cm and an axial FOV of 21 cm. The purpose of this study was to compare the performance of the two modes of HOTPET for studying small brain lesions (< 10 mm) to measure the effective gain in lesion detection by using a dedicated brain PET with higher sensitivity. We compared the two modes of camera by scanning the Hoffman brain phantom with three small lesion inserts for several lesion activity contrasts. The inner diameters of the lesion were in the range of 3.95 mm to 7.86 mm. The ratio of the activity concentration in the lesions to the activity in the background (gray matter regions) ranged from 1 to 20. Lesions were compared visually and also by measuring the contrast recovery and by studying their profiles. Visual inspection of the images showed that the 7.86-mm lesion with lesion to background activity ratio of 2 or higher was observable in both modes of the HOTPET. The 4.95-mm lesion was observable with lesion to background activity ratio of 4 (5.5) or higher in HOTPET brain (whole-body) mode. The 3.95-mm lesion was observable with lesion to background activity ratio of 8 (11) or higher in the brain (whole-body) mode. This study demonstrated the advantage of the brain mode of the HOTPET camera for detection of small lesions (<5 mm) compared to its whole-body mode and also compared to a simulated whole-body camera with crystals twice the average size of HOTPET's crystals.

Original languageEnglish (US)
Title of host publication2009 IEEE Nuclear Science Symposium Conference Record, NSS/MIC 2009
Pages2674-2678
Number of pages5
DOIs
StatePublished - 2009
Event2009 IEEE Nuclear Science Symposium Conference Record, NSS/MIC 2009 - Orlando, FL, United States
Duration: Oct 25 2009Oct 31 2009

Publication series

NameIEEE Nuclear Science Symposium Conference Record
ISSN (Print)1095-7863

Other

Other2009 IEEE Nuclear Science Symposium Conference Record, NSS/MIC 2009
Country/TerritoryUnited States
CityOrlando, FL
Period10/25/0910/31/09

ASJC Scopus subject areas

  • Radiation
  • Nuclear and High Energy Physics
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'A Hoffman brain phantom lesion study with the transformable HOTPET camera'. Together they form a unique fingerprint.

Cite this