Purpose: Task Group (TG) 224 was established by the American Association of Physicists in Medicine's Science Council under the Radiation Therapy Committee and Work Group on Particle Beams. The group was charged with developing comprehensive quality assurance (QA) guidelines and recommendations for the three commonly employed proton therapy techniques for beam delivery: scattering, uniform scanning, and pencil beam scanning. This report supplements established QA guidelines for therapy machine performance for other widely used modalities, such as photons and electrons (TG 142, TG 40, TG 24, TG 22, TG 179, and Medical Physics Practice Guideline 2a) and shares their aims of ensuring the safe, accurate, and consistent delivery of radiation therapy dose distributions to patients. Methods: To provide a basis from which machine-specific QA procedures can be developed, the report first describes the different delivery techniques and highlights the salient components of the related machine hardware. Depending on the particular machine hardware, certain procedures may be more or less important, and each institution should investigate its own situation. Results: In lieu of such investigations, this report identifies common beam parameters that are typically checked, along with the typical frequencies of those checks (daily, weekly, monthly, or annually). The rationale for choosing these checks and their frequencies is briefly described. Short descriptions of suggested tools and procedures for completing some of the periodic QA checks are also presented. Conclusion: Recommended tolerance limits for each of the recommended QA checks are tabulated, and are based on the literature and on consensus data from the clinical proton experience of the task group members. We hope that this and other reports will serve as a reference for clinical physicists wishing either to establish a proton therapy QA program or to evaluate an existing one.

Original languageEnglish (US)
JournalMedical physics
DOIs
StatePublished - Jan 1 2019

Fingerprint

Proton Therapy
Radiotherapy
Guidelines
Physics
Photons
Practice Guidelines
Protons
Medicine
Electrons

Keywords

  • QA
  • particle beams
  • proton therapy
  • quality assurance
  • radiotherapy

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

Arjomandy, B., Taylor, P., Ainsley, C., Safai, S., Sahoo, N., Pankuch, M., ... Kase, Y. (2019). AAPM task group 224: Comprehensive proton therapy machine quality assurance. Medical physics. https://doi.org/10.1002/mp.13622

AAPM task group 224 : Comprehensive proton therapy machine quality assurance. / Arjomandy, Bijan; Taylor, Paige; Ainsley, Christopher; Safai, Sairos; Sahoo, Narayan; Pankuch, Mark; Farr, Jonathan B.; Yong Park, Sung; Klein, Eric; Flanz, Jacob; Yorke, Ellen D.; Followill, David S; Kase, Yuki.

In: Medical physics, 01.01.2019.

Research output: Contribution to journalArticle

Arjomandy, B, Taylor, P, Ainsley, C, Safai, S, Sahoo, N, Pankuch, M, Farr, JB, Yong Park, S, Klein, E, Flanz, J, Yorke, ED, Followill, DS & Kase, Y 2019, 'AAPM task group 224: Comprehensive proton therapy machine quality assurance', Medical physics. https://doi.org/10.1002/mp.13622
Arjomandy, Bijan ; Taylor, Paige ; Ainsley, Christopher ; Safai, Sairos ; Sahoo, Narayan ; Pankuch, Mark ; Farr, Jonathan B. ; Yong Park, Sung ; Klein, Eric ; Flanz, Jacob ; Yorke, Ellen D. ; Followill, David S ; Kase, Yuki. / AAPM task group 224 : Comprehensive proton therapy machine quality assurance. In: Medical physics. 2019.
@article{1b175e33f9944221a330d87cee01629a,
title = "AAPM task group 224: Comprehensive proton therapy machine quality assurance",
abstract = "Purpose: Task Group (TG) 224 was established by the American Association of Physicists in Medicine's Science Council under the Radiation Therapy Committee and Work Group on Particle Beams. The group was charged with developing comprehensive quality assurance (QA) guidelines and recommendations for the three commonly employed proton therapy techniques for beam delivery: scattering, uniform scanning, and pencil beam scanning. This report supplements established QA guidelines for therapy machine performance for other widely used modalities, such as photons and electrons (TG 142, TG 40, TG 24, TG 22, TG 179, and Medical Physics Practice Guideline 2a) and shares their aims of ensuring the safe, accurate, and consistent delivery of radiation therapy dose distributions to patients. Methods: To provide a basis from which machine-specific QA procedures can be developed, the report first describes the different delivery techniques and highlights the salient components of the related machine hardware. Depending on the particular machine hardware, certain procedures may be more or less important, and each institution should investigate its own situation. Results: In lieu of such investigations, this report identifies common beam parameters that are typically checked, along with the typical frequencies of those checks (daily, weekly, monthly, or annually). The rationale for choosing these checks and their frequencies is briefly described. Short descriptions of suggested tools and procedures for completing some of the periodic QA checks are also presented. Conclusion: Recommended tolerance limits for each of the recommended QA checks are tabulated, and are based on the literature and on consensus data from the clinical proton experience of the task group members. We hope that this and other reports will serve as a reference for clinical physicists wishing either to establish a proton therapy QA program or to evaluate an existing one.",
keywords = "QA, particle beams, proton therapy, quality assurance, radiotherapy",
author = "Bijan Arjomandy and Paige Taylor and Christopher Ainsley and Sairos Safai and Narayan Sahoo and Mark Pankuch and Farr, {Jonathan B.} and {Yong Park}, Sung and Eric Klein and Jacob Flanz and Yorke, {Ellen D.} and Followill, {David S} and Yuki Kase",
year = "2019",
month = "1",
day = "1",
doi = "10.1002/mp.13622",
language = "English (US)",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",

}

TY - JOUR

T1 - AAPM task group 224

T2 - Comprehensive proton therapy machine quality assurance

AU - Arjomandy, Bijan

AU - Taylor, Paige

AU - Ainsley, Christopher

AU - Safai, Sairos

AU - Sahoo, Narayan

AU - Pankuch, Mark

AU - Farr, Jonathan B.

AU - Yong Park, Sung

AU - Klein, Eric

AU - Flanz, Jacob

AU - Yorke, Ellen D.

AU - Followill, David S

AU - Kase, Yuki

PY - 2019/1/1

Y1 - 2019/1/1

N2 - Purpose: Task Group (TG) 224 was established by the American Association of Physicists in Medicine's Science Council under the Radiation Therapy Committee and Work Group on Particle Beams. The group was charged with developing comprehensive quality assurance (QA) guidelines and recommendations for the three commonly employed proton therapy techniques for beam delivery: scattering, uniform scanning, and pencil beam scanning. This report supplements established QA guidelines for therapy machine performance for other widely used modalities, such as photons and electrons (TG 142, TG 40, TG 24, TG 22, TG 179, and Medical Physics Practice Guideline 2a) and shares their aims of ensuring the safe, accurate, and consistent delivery of radiation therapy dose distributions to patients. Methods: To provide a basis from which machine-specific QA procedures can be developed, the report first describes the different delivery techniques and highlights the salient components of the related machine hardware. Depending on the particular machine hardware, certain procedures may be more or less important, and each institution should investigate its own situation. Results: In lieu of such investigations, this report identifies common beam parameters that are typically checked, along with the typical frequencies of those checks (daily, weekly, monthly, or annually). The rationale for choosing these checks and their frequencies is briefly described. Short descriptions of suggested tools and procedures for completing some of the periodic QA checks are also presented. Conclusion: Recommended tolerance limits for each of the recommended QA checks are tabulated, and are based on the literature and on consensus data from the clinical proton experience of the task group members. We hope that this and other reports will serve as a reference for clinical physicists wishing either to establish a proton therapy QA program or to evaluate an existing one.

AB - Purpose: Task Group (TG) 224 was established by the American Association of Physicists in Medicine's Science Council under the Radiation Therapy Committee and Work Group on Particle Beams. The group was charged with developing comprehensive quality assurance (QA) guidelines and recommendations for the three commonly employed proton therapy techniques for beam delivery: scattering, uniform scanning, and pencil beam scanning. This report supplements established QA guidelines for therapy machine performance for other widely used modalities, such as photons and electrons (TG 142, TG 40, TG 24, TG 22, TG 179, and Medical Physics Practice Guideline 2a) and shares their aims of ensuring the safe, accurate, and consistent delivery of radiation therapy dose distributions to patients. Methods: To provide a basis from which machine-specific QA procedures can be developed, the report first describes the different delivery techniques and highlights the salient components of the related machine hardware. Depending on the particular machine hardware, certain procedures may be more or less important, and each institution should investigate its own situation. Results: In lieu of such investigations, this report identifies common beam parameters that are typically checked, along with the typical frequencies of those checks (daily, weekly, monthly, or annually). The rationale for choosing these checks and their frequencies is briefly described. Short descriptions of suggested tools and procedures for completing some of the periodic QA checks are also presented. Conclusion: Recommended tolerance limits for each of the recommended QA checks are tabulated, and are based on the literature and on consensus data from the clinical proton experience of the task group members. We hope that this and other reports will serve as a reference for clinical physicists wishing either to establish a proton therapy QA program or to evaluate an existing one.

KW - QA

KW - particle beams

KW - proton therapy

KW - quality assurance

KW - radiotherapy

UR - http://www.scopus.com/inward/record.url?scp=85067656009&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85067656009&partnerID=8YFLogxK

U2 - 10.1002/mp.13622

DO - 10.1002/mp.13622

M3 - Article

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

ER -