Comparison of the accuracy of US-guided biopsy of breast masses performed with 14-gauge, 16-gauge and 18-gauge automated cutting needle biopsy devices, and review of the literature

Monica L. Huang, Kenneth Hess, Rosalind P. Candelaria, Mohammad Eghtedari, Beatriz E. Adrada, Nour Sneige, Bruno D. Fornage

Research output: Contribution to journalArticlepeer-review

17 Scopus citations

Abstract

Objective: To compare the diagnostic accuracy of ultrasound (US)-guided core needle biopsy (CNB) of breast masses performed with 14-gauge, 16-gauge and 18-gauge needles. Methods: We retrospectively reviewed the charts of 1,112 patients who underwent US-guided breast CNB with 14-gauge, 16-gauge and 18-gauge needles. Cases with surgical excision or a minimum of 2 years of imaging follow-up were included. Rates of sample inadequacy, discordance with surgical or imaging findings and upgrade of DCIS to invasive cancer or high-risk lesion to in situ or invasive cancer were computed for each needle size. Results: The study included 703 CNBs: 203 performed with 14-gauge, 235 with 16-gauge and 265 with 18-gauge needles. There were no significant differences between 14-gauge, 16-gauge and 18-gauge needles in rates of specimen inadequacy (0 %, 0.4 % and 1.9 %, respectively) (p = 0.084); surgical discordance (2.6 %, 2.9 % and 3.8 %) (p = 0.76); imaging discordance (0 %, 0 % and 2 %) (p = 1.0); DCIS upgrade (43 %, 43 % and 36 %) (p = 1.00) or high-risk lesion upgrade (38 %, 25 % and 55 %) (p = 0.49). Conclusion: There was no statistically significant difference in diagnostic accuracy of US-guided CNB of breast masses performed with 14-gauge, 16-gauge and 18-gauge needles. Key Points: • Percutaneous image-guided breast core needle biopsy (CNB) is the standard of care. • Breast CNB with 14-gauge, 16-gauge and 18-gauge needles has similar diagnostic accuracy. • Smaller gauge needles can be confidently used for ultrasound-guided breast CNB.

Original languageEnglish (US)
Pages (from-to)2928-2933
Number of pages6
JournalEuropean Radiology
Volume27
Issue number7
DOIs
StatePublished - Jul 1 2017

Keywords

  • Biopsy
  • Breast
  • Core
  • Needle
  • Ultrasound

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

MD Anderson CCSG core facilities

  • Biostatistics Resource Group

Fingerprint

Dive into the research topics of 'Comparison of the accuracy of US-guided biopsy of breast masses performed with 14-gauge, 16-gauge and 18-gauge automated cutting needle biopsy devices, and review of the literature'. Together they form a unique fingerprint.

Cite this