Purpose: The goal of total scalp irradiation (TSI) is to deliver a uniform dose to the scalp, which requires the use of a bolus cap. Most current methods for fabricating bolus caps are laborious, yet still result in nonconformity and low reproducibility, which can lead to nonuniform irradiation of the scalp. We developed and validated patient-specific bolus caps for TSI using three-dimensional (3D) printing. Methods and materials: 3D-printing materials were radiologically analyzed to identify a material with properties suitable for use as a bolus cap. A Python script was developed within a commercial treatment planning system to automate the creation of a ready-to-print, patient-specific 3D bolus cap model. A bolus cap was printed for an anthropomorphic head phantom using a commercial vendor and a computed tomography simulation of the anthropomorphic head phantom and bolus cap was used to create a volumetric-modulated arc therapy TSI treatment plan. The planned treatment was delivered to the head phantom and dosimetric validation was performed using thermoluminescent dosimeters (TLD). The developed procedure was used to create a bolus cap for a clinical TSI patient, and in vivo TLD measurements were acquired for several fractions. Results: Agilus-60 was validated as a new 3D-printing material suitable for use as bolus. A 3D-printed Agilus-60 bolus cap had excellent conformality to the phantom scalp, with a maximum air gap of 4 mm. TLD measurements showed that the bolus cap generated a uniform dose to the scalp within a 2.7% standard deviation, and the delivered doses agreed with calculated doses to within 2.4% on average. The patient bolus was conformal and the average difference between TLD measured and planned doses was 5.3%. Conclusions: We have developed a workflow to 3D-print highly conformal bolus caps for TSI and demonstrated these caps can reproducibly generate a uniform dose to the scalp.

Original languageEnglish (US)
Pages (from-to)89-96
Number of pages8
JournalJournal of applied clinical medical physics
Volume20
Issue number3
DOIs
StatePublished - Mar 2019

Keywords

  • 3D printing
  • bolus
  • dosimetry
  • total scalp irradiation

ASJC Scopus subject areas

  • Radiation
  • Instrumentation
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Development and validation of a 3D-printed bolus cap for total scalp irradiation'. Together they form a unique fingerprint.

Cite this