Evaluation of an automated grid artifact detection system for quality control in digital mammography

Christopher J. MacLellan, Rick R. Layman, William Geiser, Dustin A. Gress, A. Kyle Jones

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

Purpose: Grid artifacts occur in digital mammography when synchronization between the grid assembly and generator is not achieved, including when malfunctions occur in the grid assembly or generator subsystems. Such artifacts are not explicitly monitored or evaluated by existing mammography quality control programs. In this study, we developed an automated method for quantifying the presence of grid artifacts in two-dimensional (2D) digital mammography images and assessed its utility as a supplement to existing quality control programs. Methods: Four digital mammography systems (Hologic Dimensions 3D 5000) were configured to automatically transfer 2D images to a server where the strength of the grid pattern, γmax, was quantified using a template-matching algorithm and stored in amySQL database. This analysis was performed on both American College of Radiology (ACR) phantom and clinical images. Changes in γmax were compared with image quality and service records to establish preliminary action limits for physicist intervention for each type of image. These action limits were applied around selected service events to evaluate their clinical utility. Results: All systems exhibited a gradual increase in γmax in ACR phantom images prior to having identical major components of the generator subsystem replaced, despite the absence of visible gridlines in the images. Retrospective analysis of phantom images suggested that physicists should consider AEC testing when γmax exceeds 0.050 and that clinical image quality may be affected when γmax exceeds 0.060. Eighteen of 19 visible grid artifacts were identified using a threshold γmax value of 0.065 in clinical images. Warning limits that indicate abnormal operation before visible degradation in image quality were also established. These warning limits were 0.046 and 0.041 for the 24 × 29 cm and 18 × 24 cm paddles, respectively. Specific malfunctions in the generator and grid subsystems can be detected by applying these limits. Conclusions: Automated monitoring of γmax provides useful information about the status of digital mammography units without affecting clinical operations. When used with appropriate action limits, this type of monitoring can help physicists identify specific equipment malfunctions before they would be detected by other quality control tests and before they affect clinical images.

Original languageEnglish (US)
Pages (from-to)3442-3450
Number of pages9
JournalMedical physics
Volume46
Issue number8
DOIs
StatePublished - Aug 2019

Keywords

  • anti-scatter grid
  • artifact
  • digital mammography
  • mammography
  • quality control

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Evaluation of an automated grid artifact detection system for quality control in digital mammography'. Together they form a unique fingerprint.

Cite this