Factors modulating 99mTc-MAA planar lung dosimetry for 90Y radioembolization

Benjamin P. Lopez, Armeen Mahvash, James P. Long, Marnix G.E.H. Lam, S. Cheenu Kappadath

Research output: Contribution to journalArticlepeer-review

Abstract

Purpose: To investigate the accuracy and biases of predicted lung shunt fraction (LSF) and lung dose (LD) calculations via 99mTc-macro-aggregated albumin (99mTc-MAA) planar imaging for treatment planning of 90Y-microsphere radioembolization. Methods and materials: LSFs in 52 planning and LDs in 44 treatment procedures were retrospectively calculated, in consecutive radioembolization patients over a 2 year interval, using 99mTc-MAA planar and SPECT/CT imaging. For each procedure, multiple planar LSFs and LDs were calculated using different: (1) contours, (2) views, (3) liver 99mTc-MAA shine-through compensations, and (4) lung mass estimations. The accuracy of each planar-based LSF and LD methodology was determined by calculating the median (range) absolute difference from SPECT/CT-based LSF and LD values, which have been demonstrated in phantom and patient studies to more accurately and reliably quantify the true LSF and LD values. Results: Standard-of-care LSF using geometric mean of lung and liver contours had median (range) absolute over-estimation of 4.4 percentage points (pp) (0.9 to 11.9 pp) from SPECT/CT LSF. Using anterior views only decreased LSF errors (2.4 pp median, −1.1 to +5.7 pp range). Planar LD over-estimations decreased when using single-view versus geometric-mean LSF (1.3 vs. 2.6 Gy median and 7.2 vs. 18.5 Gy maximum using 1000 g lung mass) but increased when using patient-specific versus standard-man lung mass (2.4 vs. 1.3 Gy median and 11.8 vs. 7.2 Gy maximum using single-view LSF). Conclusions: Calculating planar LSF from lung and liver contours of a single view and planar LD using that same LSF and 1000 g lung mass was found to improve accuracy and minimize bias in planar lung dosimetry.

Original languageEnglish (US)
Article numbere13734
JournalJournal of applied clinical medical physics
Volume23
Issue number12
DOIs
StatePublished - Dec 2022

Keywords

  • Y radioembolization
  • Tc-macro-aggregated albumin
  • lung dosimetry
  • planar scintigraphy

ASJC Scopus subject areas

  • Radiation
  • Instrumentation
  • Radiology Nuclear Medicine and imaging

MD Anderson CCSG core facilities

  • Biostatistics Resource Group

Fingerprint

Dive into the research topics of 'Factors modulating 99mTc-MAA planar lung dosimetry for 90Y radioembolization'. Together they form a unique fingerprint.

Cite this