Abstract
This study is aimed to introduce a novel multi-sensor-based dosimetry platform for real-time plan monitoring in HDR brachytherapy: IViST (In Vivo Source Tracking). IViST is a platform composed of three components: 1) an optimized and characterized multi-point plastic scintillator dosimeter (3 points mPSD; using BCF-60, BCF-12, and BCF-10 scintillators), 2) a compact assembly of photomultiplier tubes (PMTs) coupled to dichroic mirrors and filters for high-sensitivity scintillation light collection, and 3) a Python-based graphical user interface used for system management and signal processing. IViST can simultaneously measure dose, triangulate source position, and measure dwell time. By making 100 000 measurements/s, IViST samples enough data to quickly perform key QA/QC tasks such as identifying wrong individual dwell time, position, or interchanged transfer tubes. By using 3 co-linear sensors and planned information for an implant geometry, the platform can also triangulate source position in real-time. A clinical trial to validate this system is presently on-going using the IViST system.
| Original language | English (US) |
|---|---|
| Article number | 012024 |
| Journal | Journal of Physics: Conference Series |
| Volume | 2167 |
| Issue number | 1 |
| DOIs | |
| State | Published - Jan 24 2022 |
| Event | 11th International Conference on 3D Radiation Dosimetry, IC3DDose 2021 - Quebec City, Virtual, Canada Duration: Jun 15 2021 → … |
ASJC Scopus subject areas
- General Physics and Astronomy
Fingerprint
Dive into the research topics of 'From conception to clinical trial: IViST, the first multi-sensor-based platform for real-time in Vivo Source Tracking in HDR brachytherapy'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS