Guidelines for anti-scatter grid use in pediatric digital radiography

Shannon Fritz, A. Kyle Jones

Research output: Contribution to journalArticlepeer-review

16 Scopus citations

Abstract

Background: Pediatric radiography presents unique challenges in balancing image quality and patient dose. Removing the anti-scatter grid reduces patient dose but also reduces image contrast. The benefit of using an anti-scatter grid decreases with decreasing patient size. Objective: To determine patient thickness thresholds for anti-scatter grid use by comparing scatter-to-primary ratio for progressively thinner patients without a grid to the scatter-to-primary ratio for a standard adult patient with a grid. Materials and methods: We used Solid Water™ phantoms ranging in thickness from 7 cm to 16 cm to simulate pediatric abdomens. The scatter-to-primary ratio without a grid was measured for each thickness at 60 kVp, 70 kVp and 80 kVp for X-ray fields of view (FOV) of 378 cm2, 690 cm2 and 1,175 cm 2 using indirect digital radiography (iDR) and computed radiography (CR). We determined thresholds for anti-scatter grid use by comparing the intersection of a fit of scatter-to-primary ratio versus patient thickness with a standard adult scatter-to-primary ratio measured for a 23-cm phantom thickness at 80 kVp with an anti-scatter grid. Dose area product (DAP) was also calculated. Results: The scatter-to-primary ratio depended strongly on FOV and weakly on kVp; however DAP increased with decreasing kVp. Threshold thicknesses for grid use varied from 5 cm for a 14x17-cm FOV using iDR to 12 cm for an 8x10-cm FOV using computed radiography. Conclusions: Removing the anti-scatter grid for small patients reduces patient dose without a substantial increase in scatter-to-primary ratio when the FOV is restricted appropriately. Radiologic technologists should base anti-scatter grid use on patient thickness and FOV rather than age.

Original languageEnglish (US)
Pages (from-to)313-321
Number of pages9
JournalPediatric Radiology
Volume44
Issue number3
DOIs
StatePublished - Mar 2014

Keywords

  • Abdominal radiography
  • Anti-scatter grid
  • Digital radiography
  • Pediatric radiography
  • Radiation dose
  • Scatter-to-primary ratio

ASJC Scopus subject areas

  • Pediatrics, Perinatology, and Child Health
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Guidelines for anti-scatter grid use in pediatric digital radiography'. Together they form a unique fingerprint.

Cite this