Implications of CT noise and artifacts for quantitative 99mTc SPECT/CT imaging

K. W. Hulme, S. C. Kappadath

Research output: Contribution to journalArticlepeer-review

11 Scopus citations

Abstract

Purpose: This paper evaluates the effects of computed tomography (CT) image noise and artifacts on quantitative single-photon emission computed-tomography (SPECT) imaging, with the aim of establishing an appropriate range of CT acquisition parameters for low-dose protocols with respect to accurate SPECT attenuation correction (AC). Methods: SPECT images of two geometric and one anthropomorphic phantom were reconstructed iteratively using CT scans acquired at a range of dose levels (CTDIvol = 0.4 to 46 mGy). Resultant SPECT image quality was evaluated by comparing mean signal, background noise, and artifacts to SPECT images reconstructed using the highest dose CT for AC. Noise injection was performed on linear-attenuation (μ) maps to determine the CT noise threshold for accurate AC. Results: High levels of CT noise (σ ∼ 200-400 HU) resulted in low μ-maps noise (σ ∼ 1%-3%). Noise levels greater than ∼10% in 140 keV μ-maps were required to produce visibly perceptible increases of ∼15% in 99mTc SPECT images. These noise levels would be achieved at low CT dose levels (CTDIvol = 4 μGy) that are over 2 orders of magnitude lower than the minimum dose for diagnostic CT scanners. CT noise could also lower (bias) the expected μ values. The relative error in reconstructed SPECT signal trended linearly with the relative shift in μ. SPECT signal was, on average, underestimated in regions corresponding with beam-hardening artifacts in CT images. Any process that has the potential to change the CT number of a region by ∼100 HU (e.g., misregistration between CT images and SPECT images due to motion, the presence of contrast in CT images) could introduce errors in μ140 keV on the order of 10%, that in turn, could introduce errors on the order of ∼10% into the reconstructed 99mTc SPECT image. Conclusions: The impact of CT noise on SPECT noise was demonstrated to be negligible for clinically achievable CT parameters. Because CT dose levels that affect SPECT quantification is low (CTDIvol ∼ 4 μGy), the low dose limit for the CT exam as part of SPECT/CT will be guided by CT image quality requirements for anatomical localization and artifact reduction. A CT technique with higher kVp in combination with lower mAs is recommended when low-dose CT images are used for AC to minimize beam-hardening artifacts.

Original languageEnglish (US)
Article number042502
JournalMedical physics
Volume41
Issue number4
DOIs
StatePublished - Apr 2014

Keywords

  • SPECT/CT
  • attenuation correction
  • low-dose CT
  • radiation dose

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Implications of CT noise and artifacts for quantitative 99mTc SPECT/CT imaging'. Together they form a unique fingerprint.

Cite this