Skip to main navigation Skip to search Skip to main content

Quality assurance of clinical MRI scanners using ACR MRI phantom: Preliminary results

  • Chien Chuan Chen
  • , Yung Liang Wan
  • , Yau Yau Wai
  • , Ho Ling Liu

Research output: Contribution to journalArticlepeer-review

Abstract

Clinical magnetic resonance imaging (MRI) scanners play an important role in the diagnosis of diseases and management of patient treatment. Quality assurance (QA) of the clinical MRI scanners is mandatory to obtain optimal images in a modern hospital. In this report, the phantom test for the American College of Radiology (ACR) MRI accreditation is used as the essential part of the MRI QA protocols. Seven important assessments of MR image quality are included as follows: geometric accuracy, high-contrast resolution, slice thickness accuracy, slice position accuracy, image intensity uniformity, percent signal ghosting, and low-contrast object detectability. In addition, signal-to-noise ratio and central frequency are monitored as well. The MRI QA procedures were applied to four clinical MRI scanners in our institute twice within 3 months. According to the QA results, the service engineers were more efficient in solving scanners problems when the ACR phantom test was run.

Original languageEnglish (US)
Pages (from-to)279-284
Number of pages6
JournalJournal of Digital Imaging
Volume17
Issue number4
DOIs
StatePublished - Dec 2004
Externally publishedYes

Keywords

  • ACR
  • Image quality
  • MRI
  • Phantom
  • Quality assurance

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging
  • Computer Science Applications

Fingerprint

Dive into the research topics of 'Quality assurance of clinical MRI scanners using ACR MRI phantom: Preliminary results'. Together they form a unique fingerprint.

Cite this