Skip to main navigation Skip to search Skip to main content

Results of the prospective, randomized, multicenter clinical trial evaluating a biosynthesized cellulose graft for repair of dural defects

  • Charles L. Rosen
  • , Gary K. Steinberg
  • , Franco Demonte
  • , Johnny B. Delashaw
  • , Stephen B. Lewis
  • , Mark E. Shaffrey
  • , Khaled Aziz
  • , Johanna Hantel
  • , Frederick F. Marciano

Research output: Contribution to journal › Review article › peer-review

Abstract

Background: After intradural cranial surgery, a dural substitute is often required for dural closure. Although preferred, limitations of autograft include local availability and additional surgical site morbidity. Thus, allografts, xenografts, and synthetics are frequently used. Objective: To report 6-month results of a randomized, controlled trial of a biosynthesized cellulose (BSC) composed duraplasty device compared with commercially available dural replacements. Methods: A total of 99 patients (62 BSC; 37 control) were treated on protocol, using a 2:1 (BSC:control) blocked randomization schedule. Physical examinations were performed pre- and postoperatively within 10 days and at 1, 3, and 6 months. Magnetic resonance imaging was performed preoperatively and at 6 months. The primary study endpoint was the absence of pseudomeningocele and extracerebral fluid collection confirmed radiographically and the absence of cerebrospinal fluid fistula at 6 months. Results: At 6 months, the primary hypothesis, noninferiority of the BSC implant compared with the control group, was confirmed (P = .0206). Overall success was achieved by 96.6% of BSC and 97.1% of control patients. No significant difference was revealed between treatment groups for surgical site infection (P = 1.0000) or wound healing assessment (P ≥ .3685) outcomes, or radiologic endpoints (P ≥ .4061). Device strength and seal quality favored BSC. Conclusion: This randomized, controlled trial establishes BSC as noninferior to commercially available dural replacement devices. BSC offers a hypothetical advantage concerning prion and other infectious agent exposure; superior handling qualities are evident. Longer term data are necessary to identify limitations of BSC and its potential equivalence to the gold standard of pericranium.

Original languageEnglish (US)
Pages (from-to)1093-1103
Number of pages11
JournalNeurosurgery
Volume69
Issue number5
DOIs
StatePublished - Nov 2011
Externally publishedYes

Keywords

  • Allograft
  • Biosynthesized cellulose
  • Dura mater
  • Dural substitute
  • Duraplasty
  • Pericranium

ASJC Scopus subject areas

  • Surgery
  • Clinical Neurology

Fingerprint