Slot scanning versus anti-scatter grid in digital mammography-comparison of low contrast performance using contrast-detail measurement

Chao Jen Lai, Chris C. Shaw, William Geiser, S. Cheenu Kappadath, Xinming Liu, Tian Peng Wang, Shu Ju Tu, Mustafa C. Altunbas

Research output: Contribution to journalConference articlepeer-review

Abstract

Slot scanning imaging techniques allow for effective scatter rejection without attenuating primary x-rays. The use of these techniques should generate better image quality for the same mean glandular dose (MGD) or a similar image quality for a lower MGD as compared to imaging techniques using an anti-scatter grid. In this study, we compared a slot scanning digital mammography system (SenoScan, Fisher Imaging Systems, Denver, CO) to a full-field digital mammography (FFDM) system used in conjunction with a 5:1 anti-scatter grid (SenoGraphe 2000D, General Electric Medical Systems, Milwaukee, WI). Images of a contrast-detail phantom (University Hospital Nijmegen, The Netherlands) were reviewed to measure the contrast-detail curves for both systems. These curves were measured at 100%, 71%, 49% and 33% of the reference mean glandular dose (MGD), as determined by photo-timing, for the Fisher system and 100% for the GE system. Soft-copy reading was performed on review workstations provided by the manufacturers. The correct observation ratios (CORs) were also computed and used to compare the performance of the two systems. The results showed that, based on the contrast-detail curves, the performance of the Fisher images, acquired at 100% and 71% of the reference MGD, was comparable to the GE images at 100% of the reference MGD. The CORs for Fisher images were 0.463 and 0.444 at 100% and 71% of the reference MGD, respectively, compared to 0.453 for the GE images at 100% of the reference MGD.

Original languageEnglish (US)
Pages (from-to)734-742
Number of pages9
JournalProceedings of SPIE - The International Society for Optical Engineering
Volume5368
Issue number2
DOIs
StatePublished - 2004
EventMedical Imaging 2004: Physics of Medical Imaging - San Diego, CA, United States
Duration: Feb 15 2004Feb 17 2004

Keywords

  • CDMAM
  • Digital mammography
  • Flat panel
  • Slot scanning

ASJC Scopus subject areas

  • Electronic, Optical and Magnetic Materials
  • Condensed Matter Physics
  • Computer Science Applications
  • Applied Mathematics
  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'Slot scanning versus anti-scatter grid in digital mammography-comparison of low contrast performance using contrast-detail measurement'. Together they form a unique fingerprint.

Cite this