Abstract
In a prospective, randomized trial, the effectiveness of sterile anesthesia breathing circuits having bacterial filters in decreasing the incidence of postoperative pulmonary infection was evaluated. 293 patients undergoing major surgical procedures were assigned randomly to receive anesthetics through a sterile, disposable circuit having a bacterial filter, or a reusable circuit that had been washed and dried after previous use. The two groups were well matched for age, sex, type of surgical procedure, type of anesthetic, and history of smoking or obstructive lung disease. A physician observer, who was unaware of these assignments, reviewed the patients' postoperative status for evidence of pulmonary infection. The overall rate of infection was 3.5%; no significant difference in the rates of infection occurred between the two groups. Five (3.6%) of the 138 patients anesthetized with sterile circuits had postoperative respiratory infection, compared with 4 (2.6%) of the 155 patients anesthetized with unsterile circuits. In both groups, maximum postoperative temperature, white-blood-cell count, and the lowest values for arterial oxygen tension were similar. The authors therefore conclude that when simple hygienic measures are followed, the routine use of sterile anesthesia breathing circuits with bacterial filters does not decrease the risk of postoperative infection.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 369-372 |
| Number of pages | 4 |
| Journal | Anesthesiology |
| Volume | 54 |
| Issue number | 5 |
| DOIs | |
| State | Published - 1981 |
ASJC Scopus subject areas
- Anesthesiology and Pain Medicine
Fingerprint
Dive into the research topics of 'Sterile anesthesia breathing circuits do not prevent postoperative pulmonary infection'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS