SU‐E‐T‐102: Evaluation of the Characteristics of TLD LiF:Mg.Ti‐100 Powder: A Measure of Consistency between Multiple Batches of Powder

P. Alvarez, J. Aguirre, S. Smith, D. Followill

Research output: Contribution to journalArticlepeer-review

Abstract

Purpose: Analyze the characteristics of TLD LiF‐100 powder between multiple LiF crystal batches. Methods: The RPC used TLD LiF‐100 encapsulated powder to verify the output for photon and electron beams for 4 to 23 MV X‐ray beams and 6 to 23 MeV electron beams, respectively, from the past 15 years. During that time period, the RPC commissioned more than 15 batches of TLD powder. Commissioning of each batch of powder encompassed determining the system sensitivity (dose response), linearity, energy and fading characteristics of each batch of powder to determine the correction factors for the calculation of dose. The system sensitivity is the signal/mg per unit known dose of 60Co for each reading session. Other correction factors account for the loss of signal (fading) between the irradiation and read dates, supralinearity of the dose response and energy differences as compared to the 60Co irradiated standards. Results: More than 15 batches of TLD were commissioned to determine correction factors for the calculation of dose. The correction for fading, a characteristic of the LiF crystal, varied by ±1% between the multiple batches. The linearity correction, between 25 and 600cGy, normalized to 300cGy, showed a maximum variation of ±3% between batches. The energy correction factors, as defined for the RPC beam output audit system varied within ±1.7% (one std dev.) for the 15 batches. The system sensitivity is highly dependent on the LiF crystal grown for each batch, specific TLD reader and reading session conditions. The system sensitivity, while keeping the readers and reading sessions constant, varied by as much as 20% between batches. Conclusions: Each batch of LiF‐100 TLD powder showed variability in their powder characteristics such that calculation of dose accurately, with minimal uncertainty, requires a new commissioning. Work supported by PHS CA010953 awarded by NCI, DHHS.

Original languageEnglish (US)
Pages (from-to)3726
Number of pages1
JournalMedical physics
Volume39
Issue number6
DOIs
StatePublished - Jun 2012

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'SU‐E‐T‐102: Evaluation of the Characteristics of TLD LiF:Mg.Ti‐100 Powder: A Measure of Consistency between Multiple Batches of Powder'. Together they form a unique fingerprint.

Cite this