A new 90Y SIR-Spheres delivery kit (SIROS D-vial and shield) has been introduced with a different physical form from the legacy V-Vial kit. Here, we establish the dose calibrator settings and exposure-rate–to–activity conversion factor to assay 90Y SIR-Spheres activity in the new SIROS kit. Methods: Eight D-vials with initial 90Y activities from 1.2 to 6.6 GBq within acrylic shields were assayed with dose calibrators and exposure-rate meters until activities decayed to approximately 0.1 GBq. The dose calibrator settings resulting in the lowest median activity errors and the best-fit slope of exposure rate versus activity were identified. Results: SIROS D-vial 90Y activity can be accurately and reliably estimated directly using setting 51 3 10 on both the CRC-15R and the CRC-55tR dose calibrators (errors within 60.5%) and indirectly with an exposure-rate reading at 30 cm using conversion factor 0.664 6 0.003 GBq/(mR/h) (R2 5 0.985). Conclusion: Dose calibrator settings and exposure-rate–to–activity conversion factor for 90Y activity assays with new SIROS kit should be updated from legacy V-Vial parameters to avoid an approximately 10% underestimation.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 256-260 |
| Number of pages | 5 |
| Journal | Journal of nuclear medicine technology |
| Volume | 52 |
| Issue number | 3 |
| DOIs | |
| State | Published - Sep 1 2024 |
Keywords
- 90Y microspheres
- dose calibrator
- ionization chamber
- SIR-Spheres
- SIROS
ASJC Scopus subject areas
- General Medicine
Fingerprint
Dive into the research topics of '90Y SIR-Spheres Activity Measurement with New SIROS D-Vial Delivery Kit'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS