90Y SIR-Spheres Activity Measurement with New SIROS D-Vial Delivery Kit

Research output: Contribution to journalArticlepeer-review

Abstract

A new 90Y SIR-Spheres delivery kit (SIROS D-vial and shield) has been introduced with a different physical form from the legacy V-Vial kit. Here, we establish the dose calibrator settings and exposure-rate–to–activity conversion factor to assay 90Y SIR-Spheres activity in the new SIROS kit. Methods: Eight D-vials with initial 90Y activities from 1.2 to 6.6 GBq within acrylic shields were assayed with dose calibrators and exposure-rate meters until activities decayed to approximately 0.1 GBq. The dose calibrator settings resulting in the lowest median activity errors and the best-fit slope of exposure rate versus activity were identified. Results: SIROS D-vial 90Y activity can be accurately and reliably estimated directly using setting 51 3 10 on both the CRC-15R and the CRC-55tR dose calibrators (errors within 60.5%) and indirectly with an exposure-rate reading at 30 cm using conversion factor 0.664 6 0.003 GBq/(mR/h) (R2 5 0.985). Conclusion: Dose calibrator settings and exposure-rate–to–activity conversion factor for 90Y activity assays with new SIROS kit should be updated from legacy V-Vial parameters to avoid an approximately 10% underestimation.

Original languageEnglish (US)
Pages (from-to)256-260
Number of pages5
JournalJournal of nuclear medicine technology
Volume52
Issue number3
DOIs
StatePublished - Sep 1 2024

Keywords

  • 90Y microspheres
  • dose calibrator
  • ionization chamber
  • SIR-Spheres
  • SIROS

ASJC Scopus subject areas

  • General Medicine

Fingerprint

Dive into the research topics of '90Y SIR-Spheres Activity Measurement with New SIROS D-Vial Delivery Kit'. Together they form a unique fingerprint.

Cite this