Skip to main navigation Skip to search Skip to main content

Validation of plastic scintillation detectors for applications in low-dose-rate brachytherapy

  • François Therriault-Proulx
  • , Luc Beaulieu
  • , Sam Beddar

Research output: Contribution to journal › Article › peer-review

Abstract

Purpose To develop a plastic scintillation detector (PSD) capable of accurately measuring dose around a low-dose-rate (LDR) iodine-125 (125I) radioactive seed as a first step toward in vivo dosimetry for prostate LDR brachytherapy. Methods Using a GEANT4 based Monte Carlo code, photon energy distribution at any position around a realistic 125I source model was obtained. This energy distribution was convolved with the expected energy response from a plastic scintillator and dosimetry accuracy was evaluated. A PSD was constructed and validated in a water phantom for the entire range of clinically relevant positions around an 125I radioactive seed. Results The effect of energy dependence on dosimetry accuracy was shown to be limited, with a maximum relative difference of 1.2% from the calibration condition. A sophisticated approach to account for the energy dependence of PSDs is, therefore, not required if the detector is calibrated using the same model of radioactive seed or a geometrically similar one. The measurements were in good agreement with theoretical models for the entire clinical range. Conclusions This study shows that PSDs can be used for accurate dosimetry in real time around a single 125I seed used in LDR prostate brachytherapy and is promising for clinical applications.

Original languageEnglish (US)
Pages (from-to)903-909
Number of pages7
JournalBrachytherapy
Volume16
Issue number4
DOIs
StatePublished - Jul 2017

Keywords

  • Brachytherapy
  • Fiber
  • In vivo dosimetry
  • LDR
  • Scintillation

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Validation of plastic scintillation detectors for applications in low-dose-rate brachytherapy'. Together they form a unique fingerprint.

Cite this