A modified 3-field technique was designed with opposed cranial fields and a single spinal field encompassing the entire spinal axis. Two methods of plan verifications were performed before the first treatment. First, a system of orthogonal rulers plus the thermoplastic head holder was used to visualize the light fields at the craniospinal junction. Second, film phantom measurements were taken to visualize the gap between the fields at the level of the spinal cord. Treatment verification entailed use of a posterior-anterior (PA) portal film and placement of radiopaque wire on the inferior border of the cranial field. More rigorous verification required a custom-fabricated orthogonal film holder. The isocenter positions of both fields when they matched were recorded using a record-and-verify system. A single extended distance spinal field collimated at 42° encompassed the entire spinal neuraxis. Data were collected from 40 fractions of craniospinal irradiation (CSI). The systematic error observed for the actual daily treatments was -0.5 mm (underlap), while the stochastic error was represented by a standard deviation of 5.39 mm. Measured data across the gapped craniospinal junction with junction shifts included revealed a dose ranging from 89.3% to 108%. CSI can be performed without direct visualization of the craniospinal junction by using the verification methods described. While the use of rigorous film verification for supine technique may have reduced the systematic error, the inability to visualize the supine craniospinal junction on skin appears to have increased the stochastic error compared to published data on such errors associated with prone craniospinal irradiation.

Original languageEnglish (US)
Pages (from-to)127-131
Number of pages5
JournalMedical Dosimetry
Volume28
Issue number2
DOIs
StatePublished - 2003

Keywords

  • CT simulation
  • Craniospinal radiation therapy
  • Medulloblastoma
  • Supine positioning

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Oncology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Verification techniques and dose distribution for computed tomographic planned supine craniospinal radiation therapy'. Together they form a unique fingerprint.

Cite this