A simple test tool for checking the correspondence of the light beam and radiation field

Daniel F. Grimm, Michael T. Gillin, Robert W. Kline

Research output: Contribution to journalArticlepeer-review

Abstract

A simple test tool for checking that the correspondence between the light beam and the radiation field is within 0.2 cm has been developed for field sizes of 5 × 5 cm, 10 × 10 cm, and 20 × 20 cm. It is also possible to check the alignment of the central ray of the light beam and the central ray of the radiation field using this same device with the aid of a lead shot grid placed in the accessory tray holder. This tool has proven to be useful as part of a quality assurance program involving a number of different therapy units.

Original languageEnglish (US)
Pages (from-to)561-562
Number of pages2
JournalInternational journal of radiation oncology, biology, physics
Volume7
Issue number4
DOIs
StatePublished - Apr 1981

Keywords

  • Light beam
  • Quality assurance
  • Radiation field

ASJC Scopus subject areas

  • Radiation
  • Oncology
  • Radiology Nuclear Medicine and imaging
  • Cancer Research