Assessment of setup uncertainty in hypofractionated liver radiation therapy with a breath-hold technique using automatic image registration-based image guidance

Gye Won Choi, Yelin Suh, Prajnan Das, Joseph Herman, Emma Holliday, Eugene Koay, Albert C. Koong, Sunil Krishnan, Bruce D. Minsky, Grace L. Smith, Cullen M. Taniguchi, Sam Beddar

Research output: Contribution to journalArticlepeer-review

10 Scopus citations

Abstract

Background: Target localization in radiation therapy is affected by numerous sources of uncertainty. Despite measures to minimize the breathing motion, the treatment of hypofractionated liver radiation therapy is further challenged by residual uncertainty coming from involuntary organ motion and daily changes in the shape and location of abdominal organs. To address the residual uncertainty, clinics implement image-guided radiation therapy at varying levels of soft-tissue contrast. This study utilized the treatment records from the patients that have received hypofractionated liver radiation therapy using in-room computed tomography (CT) imaging to assess the setup uncertainty and to estimate the appropriate planning treatment volume (PTV) margins in the absence of in-room CT imaging. Methods: We collected 917 pre-treatment daily in-room CT images from 69 patients who received hypofractionated radiation therapy to the liver with the inspiration breath-hold technique. For each treatment, the daily CT was initially aligned to the planning CT based on the shape of the liver automatically using a CT-CT alignment software. After the initial alignment, manual shift corrections were determined by visual inspection of the two images, and the corrections were applied to shift the patient to the physician-approved treatment position. Considering the final alignment as the gold-standard setup, systematic and random uncertainties in the automatic alignment were quantified, and the uncertainties were used to calculate the PTV margins. Results: The median discrepancy between the final and automatic alignment was 1.1 mm (0-24.3 mm), and 38% of treated fractions required manual corrections of ≥3 mm. The systematic uncertainty was 1.5 mm in the anterior-posterior (AP) direction, 1.1 mm in the left-right (LR) direction, and 2.4 mm in the superior-inferior (SI) direction. The random uncertainty was 2.2 mm in the AP, 1.9 mm in the LR, and 2.2 mm in the SI direction. The PTV margins recommended to be used in the absence of in-room CT imaging were 5.3 mm in the AP, 3.5 mm in the LR, and 5.1 mm in the SI direction. Conclusions: Manual shift correction based on soft-tissue alignment is substantial in the treatment of the abdominal region. In-room CT can reduce PTV margin by up to 5 mm, which may be especially beneficial for dose escalation and normal tissue sparing in hypofractionated liver radiation therapy.

Original languageEnglish (US)
Article number154
JournalRadiation Oncology
Volume14
Issue number1
DOIs
StatePublished - Aug 30 2019

Keywords

  • IGRT
  • In-room CT
  • Liver radiotherapy
  • PTV margin
  • Setup uncertainty

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging

MD Anderson CCSG core facilities

  • Clinical Trials Office

Fingerprint

Dive into the research topics of 'Assessment of setup uncertainty in hypofractionated liver radiation therapy with a breath-hold technique using automatic image registration-based image guidance'. Together they form a unique fingerprint.

Cite this