Commissioning of a grid-based Boltzmann solver for cervical cancer brachytherapy treatment planning with shielded colpostats

Justin K. Mikell, Ann H. Klopp, Michael Price, Firas Mourtada

Research output: Contribution to journalArticlepeer-review

18 Scopus citations

Abstract

Purpose: We sought to commission a gynecologic shielded colpostat analytic model provided from a treatment planning system (TPS) library. We have reported retrospectively the dosimetric impact of this applicator model in a cohort of patients. Methods and Materials: A commercial TPS with a grid-based Boltzmann solver (GBBS) was commissioned for 192Ir high-dose-rate (HDR) brachytherapy for cervical cancer with stainless steel-shielded colpostats. Verification of the colpostat analytic model was verified using a radiograph and vendor schematics. MCNPX v2.6 Monte Carlo simulations were performed to compare dose distributions around the applicator in water with the TPS GBBS dose predictions. Retrospectively, the dosimetric impact was assessed over 24 cervical cancer patients' HDR plans. Results: Applicator (TPS ID #AL13122005) shield dimensions were within 0.4mm of the independent shield dimensions verification. GBBS profiles in planes bisecting the cap around the applicator agreed with Monte Carlo simulations within 2% at most locations; differing screw representations resulted in differences of up to 9%. For the retrospective study, the GBBS doses differed from TG-43 as follows (meanvalue ±standard deviation [min, max]): International Commission on Radiation units [ICRU]rectum (-8.4±2.5% [-14.1, -4.1%]), ICRUbladder (-7.2±3.6% [-15.7, -2.1%]), D2cc-rectum (-6.2±2.6% [-11.9, -0.8%]), D2cc-sigmoid (-5.6±2.6% [-9.3, -2.0%]), and D2cc-bladder (-3.4±1.9% [-7.2, -1.1%]). Conclusions: As brachytherapy TPSs implement advanced model-based dose calculations, the analytic applicator models stored in TPSs should be independently validated before clinical use. For this cohort, clinically meaningful differences (>5%) from TG-43 were observed. Accurate dosimetric modeling of shielded applicators may help to refine organ toxicity studies.

Original languageEnglish (US)
Pages (from-to)645-653
Number of pages9
JournalBrachytherapy
Volume12
Issue number6
DOIs
StatePublished - Nov 2013

Keywords

  • Acuros
  • Cervical cancer
  • Dose calculation
  • Ir
  • Shielded applicator

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging

MD Anderson CCSG core facilities

  • Clinical Trials Office

Fingerprint

Dive into the research topics of 'Commissioning of a grid-based Boltzmann solver for cervical cancer brachytherapy treatment planning with shielded colpostats'. Together they form a unique fingerprint.

Cite this