TY - GEN
T1 - Design and development of a gapless ring with modular PMT-quadrant-sharing detector (PQS) for a time-of-flight PET camera
AU - Li, Hongdi
AU - Zhang, Yuxuan
AU - Ramirez, Rocio
AU - An, Shaohui
AU - Wang, Chao
AU - Liu, Shitao
AU - Dong, Yun
AU - Baghaei, Hossain
AU - Wong, Wai Hoi
N1 - Copyright:
Copyright 2014 Elsevier B.V., All rights reserved.
PY - 2013
Y1 - 2013
N2 - We have successfully developed a new high-resolution PMT-Quadrant-Sharing (PQS)[1] PET panel detector system. Each bank consists of 48 × 112 LYSO detector elements with a pixel size of 2.35 × 2.35 × 15.5 mm 3. After grinding a regular detector bank into a half pentagon shape with tapered-end crystals, 24 half pentagon banks will form a gapless cylindrical ring with 87-cm in diameter and 27.6-cm in axial FOV. Two gapless adjacent panels share one row of PMT to extend the PQS configuration beyond one panel for maximizing the PMT usage economically. The entire ring has 504 PQS blocks consisted of 129,024 detectors and coupled to 576 PMTs (φ 38-mm). The aim of this study is to achieve an optimized PET system performance in spatial resolution, sensitivity, emission-scan duration, detector time-resolution, electronic simplicity, count-rate, manufacturing and service cost.
AB - We have successfully developed a new high-resolution PMT-Quadrant-Sharing (PQS)[1] PET panel detector system. Each bank consists of 48 × 112 LYSO detector elements with a pixel size of 2.35 × 2.35 × 15.5 mm 3. After grinding a regular detector bank into a half pentagon shape with tapered-end crystals, 24 half pentagon banks will form a gapless cylindrical ring with 87-cm in diameter and 27.6-cm in axial FOV. Two gapless adjacent panels share one row of PMT to extend the PQS configuration beyond one panel for maximizing the PMT usage economically. The entire ring has 504 PQS blocks consisted of 129,024 detectors and coupled to 576 PMTs (φ 38-mm). The aim of this study is to achieve an optimized PET system performance in spatial resolution, sensitivity, emission-scan duration, detector time-resolution, electronic simplicity, count-rate, manufacturing and service cost.
KW - PMT-Quadrant-Sharing
KW - Positron emission tomography (PET)
KW - TOF
KW - high-resolution
KW - high-sensitivity
UR - http://www.scopus.com/inward/record.url?scp=84904167928&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84904167928&partnerID=8YFLogxK
U2 - 10.1109/NSSMIC.2013.6829073
DO - 10.1109/NSSMIC.2013.6829073
M3 - Conference contribution
AN - SCOPUS:84904167928
SN - 9781479905348
T3 - IEEE Nuclear Science Symposium Conference Record
BT - 2013 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2013
PB - Institute of Electrical and Electronics Engineers Inc.
T2 - 2013 60th IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2013
Y2 - 27 October 2013 through 2 November 2013
ER -