T1 - Dose perturbations from implanted helical gold markers in proton therapy of prostate cancer
AU - Giebeler, Annelise
AU - Fontenot, Jonas
AU - Balter, Peter
AU - Ciangaru, George
AU - Zhu, Ronald
AU - Newhauser, Wayne
PY - 2009
Y1 - 2009
N2 - Implanted gold fiducial markers are widely used in radiation therapy to improve targeting accuracy. Recent investigations have revealed that metallic fiducial markers can cause severe perturbations in dose distributions for proton therapy, suggesting smaller markers should be considered. The objective of this study was to estimate the dosimetric impact of small gold markers in patients receiving proton therapy for prostate cancer. Small, medium, and large helical wire markers with lengths of 10 mm and helix diameters of 0.35 mm, 0.75 mm, and 1.15 mm, respectively, were implanted in an anthropomorphic phantom. Radiographic visibility was confirmed using a kilovoltage x-ray imaging system, and dose perturbations were predicted from Monte Carlo simulations and confirmed by measurements. Monte Carlo simulations indicated that size of dose perturbation depended on marker size, orientation, and distance from the beam's end of range. Specifically, the perturbation of proton dose for the lateral-opposed-pair treatment technique was 31% for large markers and 23% for medium markers in a typical oblique orientation. Results for perpendicular and parallel orientations were respectively lower and higher. Consequently, these markers are not well suited for use in patients receiving proton therapy for prostate cancer. Dose perturbation was not observed for the small markers, but these markers were deemed too fragile for transrectal implantation in the prostate.
AB - Implanted gold fiducial markers are widely used in radiation therapy to improve targeting accuracy. Recent investigations have revealed that metallic fiducial markers can cause severe perturbations in dose distributions for proton therapy, suggesting smaller markers should be considered. The objective of this study was to estimate the dosimetric impact of small gold markers in patients receiving proton therapy for prostate cancer. Small, medium, and large helical wire markers with lengths of 10 mm and helix diameters of 0.35 mm, 0.75 mm, and 1.15 mm, respectively, were implanted in an anthropomorphic phantom. Radiographic visibility was confirmed using a kilovoltage x-ray imaging system, and dose perturbations were predicted from Monte Carlo simulations and confirmed by measurements. Monte Carlo simulations indicated that size of dose perturbation depended on marker size, orientation, and distance from the beam's end of range. Specifically, the perturbation of proton dose for the lateral-opposed-pair treatment technique was 31% for large markers and 23% for medium markers in a typical oblique orientation. Results for perpendicular and parallel orientations were respectively lower and higher. Consequently, these markers are not well suited for use in patients receiving proton therapy for prostate cancer. Dose perturbation was not observed for the small markers, but these markers were deemed too fragile for transrectal implantation in the prostate.
KW - Fiducial markers
KW - Heavier particle dosimetry
KW - Neutron
KW - Proton
UR - http://www.scopus.com/inward/record.url?scp=70049117942&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=70049117942&partnerID=8YFLogxK
U2 - 10.1120/jacmp.v10i1.2875
DO - 10.1120/jacmp.v10i1.2875
M3 - Article
C2 - 19223836
AN - SCOPUS:70049117942
SN - 1526-9914
VL - 10
SP - 63
EP - 70
JO - Journal of applied clinical medical physics
JF - Journal of applied clinical medical physics
IS - 1
ER -