T1 - Dosimetric comparison of craniospinal irradiation using different tomotherapy techniques
AU - Zhang, Xin
AU - Penagaricano, J.
AU - Han, E. Y.
AU - Morrill, S.
AU - Hardee, M.
AU - Liang, X.
AU - Gupta, S. K.
AU - Corry, P. M.
AU - Ratanatharathom, V.
N1 - Publisher Copyright:
© The Author(s) 2014.
PY - 2015/1/1
Y1 - 2015/1/1
N2 - The objective of this study is to compare the new and conventional tomotherapy treatment techniques and to evaluate dosimetric differences between them. A dosimetric analysis was performed by comparing planning target volume (PTV) median dose, 95% of PTV dose coverage, Paddick conformity index (CI), homogeneity index (HI), whole-body integral dose, and OAR median doses. The beam on time (BOT) and the effect of different jaw sizes and pitch values was studied. The study results indicated that the PTV dose coverage for all the techniques was comparable. Treatment plans using dynamic jaw reduced OAR doses to structures located at the treatment field edge compared to fixed jaw plans. The HT-3DCRT plans resulted in higher OAR doses to kidney, liver, and lung compared to the other techniques, and TD-IMRT provided the best dose sparing to liver compared to other techniques. Whole-body integral dose differences were found to be insignificant among the techniques. BOT was found to be higher for fixed jaw treatment plan compared to dynamic jaw plan and comparable between all treatment techniques with 5-cm dynamic jaw. In studying effect of jaw size, better OAR sparing and HI were found for 2.5-cm jaw but at the expense of doubling of BOT as compared to 5-cm jaw. There was no significant improvement found in OAR sparing when the pitch value was increased. Increasing the pitch from 0.2 to 0.43, the CI was improved, HI improved only for 5-cm jaw size, and BOT decreased to approximately half of its original time.
AB - The objective of this study is to compare the new and conventional tomotherapy treatment techniques and to evaluate dosimetric differences between them. A dosimetric analysis was performed by comparing planning target volume (PTV) median dose, 95% of PTV dose coverage, Paddick conformity index (CI), homogeneity index (HI), whole-body integral dose, and OAR median doses. The beam on time (BOT) and the effect of different jaw sizes and pitch values was studied. The study results indicated that the PTV dose coverage for all the techniques was comparable. Treatment plans using dynamic jaw reduced OAR doses to structures located at the treatment field edge compared to fixed jaw plans. The HT-3DCRT plans resulted in higher OAR doses to kidney, liver, and lung compared to the other techniques, and TD-IMRT provided the best dose sparing to liver compared to other techniques. Whole-body integral dose differences were found to be insignificant among the techniques. BOT was found to be higher for fixed jaw treatment plan compared to dynamic jaw plan and comparable between all treatment techniques with 5-cm dynamic jaw. In studying effect of jaw size, better OAR sparing and HI were found for 2.5-cm jaw but at the expense of doubling of BOT as compared to 5-cm jaw. There was no significant improvement found in OAR sparing when the pitch value was increased. Increasing the pitch from 0.2 to 0.43, the CI was improved, HI improved only for 5-cm jaw size, and BOT decreased to approximately half of its original time.
KW - Dosimetry
KW - Dynamic jaw
KW - Helical tomotherapy 3DCRT
KW - Helical tomotherapy IMRT
KW - TomoDirect IMRT
UR - http://www.scopus.com/inward/record.url?scp=84960376484&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84960376484&partnerID=8YFLogxK
U2 - 10.1177/1533034614556932
DO - 10.1177/1533034614556932
M3 - Article
C2 - 25398680
AN - SCOPUS:84960376484
SN - 1533-0346
VL - 14
SP - 440
EP - 446
JO - Technology in Cancer Research and Treatment
JF - Technology in Cancer Research and Treatment
IS - 4
ER -