Dosimetric validation of the Gamma Knife® IconTM plan adaptation and high-definition motion management system with a motorized anthropomorphic head phantom

Eun Young Han, Dershan Luo, Jong Oh Kim, Kelly Tharp, Zhifei Wen, Tina M. Briere

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

Purpose: To perform dosimetric validation of the plan adaptation and high-definition motion management (HDMM) system of Gamma Knife® IconTM in various clinical scenarios. Methods and materials: We built an assembly for a pitch-adjustable anthropomorphic head phantom. We then used films to measure dosimetric and positional accuracy in 13 clinical scenarios, including movement near HDMM thresholds, multiple plan adaptations, frequent coughing, and initial setup error. Results: The dose for the superiorly located 4-mm shot was decreased up to 7-13% near 2- to 3-mm HDMM thresholds in the chin-down position. Dosimetric deviation was within ±3.5% for initial pitch angles of up to 20°. Multiple treatment interruption and frequent coughing did not cause substantial dosimetric deviation (<2%). Conclusion: Our results indicated that dosimetric accuracy of the Gamma Knife® IconTM system is reliable even in extreme treatment conditions. However, the user should exercise caution for superiorly located small lesions with an HDMM threshold ≥2 mm or in the scenario of large initial setup error.

Original languageEnglish (US)
Pages (from-to)217-226
Number of pages10
JournalJournal of Radiosurgery and SBRT
Volume6
Issue number3
StatePublished - 2019

Keywords

  • Gamma Knife Icon
  • HDMM
  • Plan adaptation

ASJC Scopus subject areas

  • Surgery
  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Dosimetric validation of the Gamma Knife® IconTM plan adaptation and high-definition motion management system with a motorized anthropomorphic head phantom'. Together they form a unique fingerprint.

Cite this