Effect of anisotropy corrections on the dynamic dose calculations in high dose rate (HDR) brachytherapy

Sang Hyun Cho, R. Muller-Runkel

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

An analytic formalism utilizing a convolution technique is presented for the incorporation of anisotropy corrections into the dynamic (or transit) dose calculations in high dose rate (HDR) brachytherapy. A simple numerical implementation of the suggested formalism is also presented. Two calculational examples are provided in order to show some effects of anisotropy corrections on the dynamic dose calculations. Based on the results from these examples and published literature, the absence of anisotropy corrections in the dynamic dose calculations could result in a dosimetric error up to the order of 0.1 Gy during the full course of HDR brachytherapy treatments depending on fractionation and geometrical arrangement. Therefore, the incorporation of anisotropy corrections into dynamic dose calculations could eliminate this kind of error in HDR brachytherapy.

Original languageEnglish (US)
Article number009
Pages (from-to)1181-1188
Number of pages8
JournalPhysics in medicine and biology
Volume39
Issue number7
DOIs
StatePublished - 1994
Externally publishedYes

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Effect of anisotropy corrections on the dynamic dose calculations in high dose rate (HDR) brachytherapy'. Together they form a unique fingerprint.

Cite this