Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies

Bijan Arjomandy, Ramesh Tailor, Aman Anand, Narayan Sahoo, Michael Gillin, Karl Prado, Milos Vicic

Research output: Contribution to journalArticlepeer-review

223 Scopus citations

Abstract

Purpose: Since the Gafchromic film EBT has been recently replaced by the newer model EBT2, its characterization, especially energy dependence, has become critically important. The energy dependence of the dose response of Gafchromic EBT2 film is evaluated for a broad range of energies from different radiation sources used in radiation therapy. Methods: The beams used for this study comprised of kilovoltage x rays (75, 125, and 250 kVp), 137Cs gamma (662 KeV), C60 o gamma (1.17-1.33 MeV), megavoltage x rays (6 and 18 MV), electron beams (6 and 20 MeV), and proton beams (100 and 250 MeV). The film's response to each of the above energies was measured over the dose range of 0.4-10 Gy, which corresponds to optical densities ranging from 0.05 to 0.74 for the film reader used. Results: The energy dependence of EBT2 was found to be relatively small within measurement uncertainties (1σ=±4.5%) for all energies and modalities. Conclusion: For relative and absolute dosimetry of radiation therapy beams, the weak energy dependence of the EBT2 makes it most suitable for clinical use compared to other films.

Original languageEnglish (US)
Pages (from-to)1942-1947
Number of pages6
JournalMedical physics
Volume37
Issue number5
DOIs
StatePublished - May 2010

Keywords

  • Densitometry
  • Dosimetry
  • EBT2
  • Energy dependence
  • Gafchromic films

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies'. Together they form a unique fingerprint.

Cite this