Film dosimetry of small electron beams for routine radiotherapy planning

Azam Niroomand rad, Michael T. Gillin, Robert W. Kline, Daniel F. Grimm

Research output: Contribution to journalLetterpeer-review

18 Scopus citations

Abstract

The characteristics of very small fields, 1–1 and 2–2 cm, of electron beams of nominal energies, 5, 7, 10, 12, 15, and 18 MeV have been studied and compared to a 10–10 cm field. A parallel plate ion chamber and film have been used to obtain various dose parameters. The central axis depth dose measurements, field flatness, uniformity index, and relative output factors are presented. It was found that satisfactory results for determining the relative output factor can be obtained from film data using a scanning densitometer. It is our conclusion that film dosimetry is acceptable in determining the necessary clinical parameters needed to treat patients with fields as small as 2–2 cm. For the 1–1 cm field size and for the electron energies greater than 10 MeV, there was substantial disagreement between the ion chamber and film data in the buildup region as well as the regions beyond the depth of maximum dose to the depth of 90% dose.

Original languageEnglish (US)
Pages (from-to)416-421
Number of pages6
JournalMedical physics
Volume13
Issue number3
DOIs
StatePublished - May 1986

Keywords

  • 87.53.10.b
  • 87.55.1
  • DEPTH DOSE DISTRIBUTIONS
  • ELECTRON BEAMS
  • ELECTRON DOSIMETRY
  • FILM DOSIMETRY
  • MEV RANGE 01-10
  • PLANNING
  • RADIOTHERAPY

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Film dosimetry of small electron beams for routine radiotherapy planning'. Together they form a unique fingerprint.

Cite this