Hemoccult stool tests. False negative results due to storage of specimens

J. R. Stroehlein, V. F. Fairbanks, V. L.W. Go, W. F. Taylor, J. H. Thompson

Research output: Contribution to journalArticlepeer-review

43 Scopus citations

Abstract

Stool samples mixed with 0 to 38 mg of hemoglobin per gram of stool were tested for occult blood by the Hemoccult (HO) method when fresh and after storage for 1 to 14 days. The intensity of the HO reaction in fresh stool samples was directly related to the hemoglobin concentration, and approximately 8 mg/g was required for a positive reaction. After samples had been stored on HO cards for 4 days, the intensity of the HO reaction was decidedly less in all. Except at the highest concentrations studied, all became HO negative or trace within 8 days. In contrast, samples of most concentrations that were stored in tightly capped containers prior to preparation of HO cards maintained a positive reaction intensity for 1 week. Nevertheless, the intensity of the HO reaction decreased significantly (P<0.01) with each method of storage. No false positives were induced by storage during this portion of the authors' study. Samples from 100 stool specimens sent to the clinical laboratory for determination of occult blood were also smeared on separate HO cards for testing immediately and after 4, 7, or 10 days of storage. Only one positive HO reaction was observed on an HO card prepared from a specimen initially classified as HO negative. In contrast, up to 40% of the stored HO cards prepared from specimens initially classified as HO positive were HO negative when analyzed 4 days later.

Original languageEnglish (US)
Pages (from-to)548-552
Number of pages5
JournalMayo Clinic Proceedings
Volume51
Issue number9
StatePublished - 1976

ASJC Scopus subject areas

  • General Medicine

Fingerprint

Dive into the research topics of 'Hemoccult stool tests. False negative results due to storage of specimens'. Together they form a unique fingerprint.

Cite this