Impact of low injected activity on data driven respiratory gating for PET/CT imaging with continuous bed motion

Joseph G. Meier, Radwan H. Diab, Trevor M. Connor, Osama R. Mawlawi

Research output: Contribution to journalArticlepeer-review

Abstract

Data driven respiratory gating (DDG) in positron emission tomography (PET) imaging extracts respiratory waveforms from the acquired PET data obviating the need for dedicated external devices. DDG performance, however, degrades with decreasing detected number of coincidence counts. In this paper, we assess the clinical impact of reducing injected activity on a new DDG algorithm designed for PET data acquired with continuous bed motion (CBM_DDG) by evaluating CBM_DDG waveforms, tumor quantification, and physician's perception of motion blur in resultant images. Forty patients were imaged on a Siemens mCT scanner in CBM mode. Reduced injected activity was simulated by generating list mode datasets with 50% and 25% of the original data (100%). CBM_DDG waveforms were compared to that of the original data over the range between the aortic arch and the center of the right kidney using the Pearson correlation coefficient (PCC). Tumor quantification was assessed by comparing the maximum standardized uptake value (SUVmax) and peak SUV (SUVpeak) of reconstructed images from the various list mode datasets using elastic motion deblurring (EMDB) reconstruction. Perceived motion blur was assessed by three radiologists of one lesion per patient on a continuous scale from no motion blur (0) to significant motion blur (3). The mean PCC of the 50% and 25% dataset waveforms was 0.74 ± 0.18 and 0.59 ± 0.25, respectively. In comparison to the 100% datasets, the mean SUVmax increased by 2.25% (p = 0.11) for the 50% datasets and by 3.91% (p = 0.16) for the 25% datasets, while SUVpeak changes were within ±0.25%. Radiologist evaluations of motion blur showed negligible changes with average values of 0.21, 0.3, and 0.28 for the 100%, 50%, and 25% datasets. Decreased injected activities degrades the resultant CBM_DDG respiratory waveforms; however this decrease has minimal impact on quantification and perceived image motion blur.

Original languageEnglish (US)
Article numbere13619
JournalJournal of applied clinical medical physics
Volume23
Issue number5
DOIs
StatePublished - May 2022

Keywords

  • low injected activity SUV quantification
  • PET/CT respiratory motion correction data driven gating

ASJC Scopus subject areas

  • Radiation
  • Instrumentation
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Impact of low injected activity on data driven respiratory gating for PET/CT imaging with continuous bed motion'. Together they form a unique fingerprint.

Cite this