Microcalcification detectability for four mammographic detectors: Flat-panel, CCD, CR, and screen/film

Xiujiang J. Rong, Chris C. Shaw, Dennis A. Johnston, Michael R. Lemacks, Xinming Liu, Gary J. Whitman, Mark J. Dryden, Tanya W. Stephens, Stephen K. Thompson, Kerry T. Krugh, Chao Jen Lai

Research output: Contribution to journalArticlepeer-review

55 Scopus citations

Abstract

Amorphous silicon/cesium iodide (a-Si:H/CsI:Tl) flat-panel (FP)-based full-field digital mammography systems have recently become commercially available for clinical use. Some investigations on physical properties and imaging characteristics of these types of detectors have been conducted and reported. In this perception study, a phantom containing simulated microcalcifications (μCs) of various sizes was imaged with four detector systems: a FP system, a small field-of-view charge coupled device (CCD) system, a high resolution computed radiography (CR) system, and a conventional mammography screen/film (SF) system. The images were reviewed by mammographers as well as nonradiologist participants. Scores reflecting confidence ratings were given and recorded for each detection task. The results were used to determine the average confidence-rating scores for the four imaging systems. Receiver operating characteristics (ROC) analysis was also performed to evaluate and compare the overall detection accuracy for the four detector systems. For calcifications of 125-140 μm in size, the FP system was found to have the best performance with the highest confidence-rating scores and the greatest detection accuracy (Az = 0.9) in the ROC analysis. The SF system was ranked second while the CCD system outperformed the CR system. The p values obtained by applying a Student t-test to the results of the ROC analysis indicate that the differences between any two systems are statistically significant (p<0.005). Differences in μC detectability for the large (150-160 μm) and small (112-125 μm) size μC groups showed a wider range of p values (not all p values are smaller than 0.005, ranging from 0.6 to <0.001) compared to the p values obtained for the medium (125-140 μm) size μC group. Using the p values to assess the statistical significance, the use of the average confidence-rating scores was not as significant as the use of the ROC analysis p value for p value.

Original languageEnglish (US)
Pages (from-to)2052-2061
Number of pages10
JournalMedical physics
Volume29
Issue number9
DOIs
StatePublished - Sep 2002

Keywords

  • Charge coupled device
  • Computed radiography
  • Flat-panel
  • Mammography
  • Microcalcification phantom
  • Screen/film