Minimally invasive electro-magnetic navigational bronchoscopy-integrated near-infrared-guided sentinel lymph node mapping in the porcine lung

Hironobu Wada, Kentaro Hirohashi, Takashi Anayama, Takahiro Nakajima, Tatsuya Kato, Harley H.L. Chan, Jimmy Qiu, Michael Daly, Robert Weersink, David A. Jaffray, Jonathan C. Irish, Thomas K. Waddell, Shaf Keshavjee, Ichiro Yoshino, Kazuhiro Yasufuku

Research output: Contribution to journalArticlepeer-review

9 Scopus citations

Abstract

Background: The use of near-infrared (NIR) fluorescence imaging with indocyanine green (ICG) for sentinel lymph node (SN) mapping has been investigated in lung cancer; however, this has not been fully adapted for minimally invasive surgery (MIS). The aim of our study was to develop a minimally invasive SN mapping integrating pre-operative electro-magnetic navigational bronchoscopy (ENB)-guided transbronchial ICG injection and intraoperative NIR thoracoscopic imaging. Methods: A NIR thoracoscope was used to visualize ICG fluorescence. ICG solutions in a 96-well plate and ex vivo porcine lungs were examined to optimize ICG concentrations and injection volumes. Transbronchial ICG injection (n=4) was assessed in comparison to a traditional transpleural approach (n=3), where after thoracotomy an ICG solution (100μL at 100μg/mL) was injected into the porcine right upper lobe for SN identification. For further translation into clinical use, transbronchial ICG injection prior to thoracotomy followed by NIR thoracoscopic imaging was validated (n=3). ENB was used for accurate targeting in two pigs with a pseudo-tumor. Results: The ICG fluorescence at 10 μg/mL was the brightest among various concentrations, unchanged by the distance between the thoracoscope and ICG solutions. Injected ICG of no more than 500μL showed a localized fluorescence area. All 7 pigs showed a bright paratracheal lymph node within 15 minutes post-injection, with persistent fluorescence for 60 minutes. The antecedent transbronchial ICG injection succeeded in SN identification in all 3 cases at the first thoracoscopic inspection within 20 minutes post-injection. The ENB system allowed accurate ICG injection surrounding the pseudo-tumors. Conclusions: ENB-guided ICG injection followed by NIR thoracoscopy was technically feasible for SN mapping in the porcine lung. This promising platform may be translated into human clinical trials and is suited for MIS.

Original languageEnglish (US)
Article numbere0126945
JournalPloS one
Volume10
Issue number5
DOIs
StatePublished - May 20 2015
Externally publishedYes

ASJC Scopus subject areas

  • General

Fingerprint

Dive into the research topics of 'Minimally invasive electro-magnetic navigational bronchoscopy-integrated near-infrared-guided sentinel lymph node mapping in the porcine lung'. Together they form a unique fingerprint.

Cite this