Multiple-bandwidth photoacoustic tomography

Geng Ku, Xueding Wang, George Stoica, Lihong V. Wang

Research output: Contribution to journalArticlepeer-review

203 Scopus citations

Abstract

Photoacoustic tomography, also referred to as optoacoustic tomography, employs short laser pulses to generate ultrasonic waves in biological tissues. The reconstructed images can be characterized by the convolution of the structure of samples, the laser pulse and the impulse response of the ultrasonic transducer used for detection. Although the laser-induced ultrasonic waves cover a wide spectral range, a single transducer can receive only part of the spectrum because of its limited bandwidth. To systematically analyse this problem, we constructed a photoacoustic tomographic system that uses multiple ultrasonic transducers simultaneously, each at a different central frequency. The photoacoustic images associated with the different transducers were compared and analysed. The system was tested by imaging both mouse brains and phantom samples. The vascular vessels in the brain were revealed by all of the transducers, but the image resolutions differed. The higher frequency detectors provided better image resolution while the lower frequency detectors delineated the major structural traits with a higher signal-noise ratio.

Original languageEnglish (US)
Pages (from-to)1329-1338
Number of pages10
JournalPhysics in medicine and biology
Volume49
Issue number7
DOIs
StatePublished - Apr 7 2004

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Multiple-bandwidth photoacoustic tomography'. Together they form a unique fingerprint.

Cite this