Performance characteristics of the 5-ring GE Discovery MI PET/CT scanner using AAPM TG-126 report

Refaat Y. AlMazrou, Shadei F. Alanazi, Mohammed H. Alzaid, Razan S. Al-Fakhranee, Shanli Ding, Osama R. Mawlawi

Research output: Contribution to journalArticlepeer-review

Abstract

Aim: To report on the performance characteristics of the 5-ring GE Discovery MI PET/CT systems using the AAPM TG-126 report and compare these results to NEMA NU 2-2012 where applicable. Materials and Methods: TG-126 testing was performed on two GE 5-Rings Discovery MI scanners. Tests performed included spatial resolution, PET/CT image-registration accuracy, sensitivity, count rate performance, accuracy of corrections, image contrast, scatter/attenuation correction, and image uniformity. All acquired data were analyzed using scanner console or free software tools as described by TG-126 and the results were then compared to published NEMA NU 2-2012 values. Results: Both scanners gave similar resolution results for TG-126 and NEMA NU 2-2012 and were within manufacturer specifications. Image-registration accuracy between PET and CT using our clinical protocol showed excellent results with values ≤1 mm. Sensitivity using TG-126 was 19.43 cps/kBq while for NEMA the value was 20.73 cps/kBq. The peak noise-equivalent counting rate was 2174 kcps at 63.1 kBq/mL and is not comparable to NEMA NU 2-2012 due to differences in phantoms and methods used to measure and calculate this parameter. The accuracy of corrections for count losses for TG-126 were expressed in SUV values and found to be within 10% of the expected SUV measurement of 1. Image contrast and scatter/attenuation correction using the TG-126 method gave acceptable results. Image uniformity assessment resulted in values within the recommended ± 5% limits. Conclusion: These results show that the 5-ring GE Discovery MI PET/CT scanner testing using TG-126 is reproducible and has similar results to NEMA NU 2-2012 tests where applicable. We hope these results start to form the basis to compare PET/CT systems using TG-126.

Original languageEnglish (US)
Article numbere14315
JournalJournal of applied clinical medical physics
Volume25
Issue number4
DOIs
StatePublished - Apr 2024

Keywords

  • AAPM TG-126
  • acceptance testing
  • NEMA NU 2-2012
  • PET/CT

ASJC Scopus subject areas

  • Radiation
  • Instrumentation
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Performance characteristics of the 5-ring GE Discovery MI PET/CT scanner using AAPM TG-126 report'. Together they form a unique fingerprint.

Cite this