Abstract
The physics quality assurance (QA) program for pencil beam scanning at the Proton Therapy Center in Houston is presented. This program consists of machine-specific and patient-specific measurements. The program follows national and international recommendations. Machine-specific QA assures the accuracy and constancy of nine different delivery components. Recommendations for daily, monthly, and annual measurements are presented. Patient-specific QA includes plan review by the physicist and verification of plan parameters and image transfer to the record-and-verify system, as well as comparison of the planned dose distribution of every treatment field with the measured dose distribution in a water-equivalent phantom. Finally, the QA programs at other institutions are briefly reviewed.
Original language | English (US) |
---|---|
Title of host publication | Proton Therapy |
Subtitle of host publication | Indications, Techniques and Outcomes |
Publisher | Elsevier |
Pages | 80-105.e2 |
ISBN (Electronic) | 9780323733496 |
ISBN (Print) | 9780323733502 |
DOIs | |
State | Published - Jan 1 2020 |
Keywords
- dosimetric QA
- machine QA
- mechanical QA
- patient-specific QA
ASJC Scopus subject areas
- General Medicine