A methodology is presented for placing deep brain stimulator electrodes under direct MR image guidance. The technique utilized a small, skull-mounted trajectory guide that is optimized for accurate alignment under MR fluoroscopy. Iterative confirmation scans are used to monitor device alignment and brain penetration. The methodology was initially tested in a human skull phantom and proved capable of achieving submillimeter accuracy over a set of 16 separate targets that were accessed. The maximum error that was obtained in this preliminary test was 2 mm, motivating use of the technique in a clinical study. Subsequently, a total of eight deep brain stimulation electrodes were placed in five patients. Satisfactory placement was achieved on the first pass in seven of eight electrodes, while two passes were required with one electrode. Mean error from the intended target on the first pass was 1.0 ± 0.8 mm (range = 0.1-1.9 mm). All procedures were considered technical successes and there were no intraoperative complications; however, one patient did develop a postoperative infection.

Original languageEnglish (US)
Pages (from-to)1107-1114
Number of pages8
JournalMagnetic resonance in medicine
Volume54
Issue number5
DOIs
StatePublished - Nov 2005
Externally publishedYes

Keywords

  • DBS electrodes
  • Implants
  • Interventional MR
  • MR fluoroscopy
  • Parkinson's disease

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Placement of deep brain stimulator electrodes using real-time high-field interventional magnetic resonance imaging'. Together they form a unique fingerprint.

Cite this