Radiation Therapy Deficiencies Identified During On-Site Dosimetry Visits by the Imaging and Radiation Oncology Core Houston Quality Assurance Center

Stephen F. Kry, Lainy Dromgoole, Paola Alvarez, Jessica Leif, Andrea Molineu, Paige Taylor, David S. Followill

Research output: Contribution to journalArticlepeer-review

18 Scopus citations

Abstract

Purpose To review the dosimetric, mechanical, and programmatic deficiencies most frequently observed during on-site visits of radiation therapy facilities by the Imaging and Radiation Oncology Core Quality Assurance Center in Houston (IROC Houston). Methods and Materials The findings of IROC Houston between 2000 and 2014, including 409 institutions and 1020 linear accelerators (linacs), were compiled. On-site evaluations by IROC Houston include verification of absolute calibration (tolerance of ±3%), relative dosimetric review (tolerances of ±2% between treatment planning system [TPS] calculation and measurement), mechanical evaluation (including multileaf collimator and kilovoltage-megavoltage isocenter evaluation against Task Group [TG]-142 tolerances), and general programmatic review (including institutional quality assurance program vs TG-40 and TG-142). Results An average of 3.1 deficiencies was identified at each institution visited, a number that has decreased slightly with time. The most common errors are tabulated and include TG-40/TG-142 compliance (82% of institutions were deficient), small field size output factors (59% of institutions had errors ≥3%), and wedge factors (33% of institutions had errors ≥3%). Dosimetric errors of ≥10%, including in beam calibration, were seen at many institutions. Conclusions There is substantial room for improvement of both dosimetric and programmatic issues in radiation therapy, which should be a high priority for the medical physics community. Particularly relevant was suboptimal beam modeling in the TPS and a corresponding failure to detect these errors by not including TPS data in the linac quality assurance process.

Original languageEnglish (US)
Pages (from-to)1094-1100
Number of pages7
JournalInternational Journal of Radiation Oncology Biology Physics
Volume99
Issue number5
DOIs
StatePublished - Dec 1 2017

ASJC Scopus subject areas

  • Radiation
  • Oncology
  • Radiology Nuclear Medicine and imaging
  • Cancer Research

Fingerprint

Dive into the research topics of 'Radiation Therapy Deficiencies Identified During On-Site Dosimetry Visits by the Imaging and Radiation Oncology Core Houston Quality Assurance Center'. Together they form a unique fingerprint.

Cite this