Receiver operating characteristic analysis for the detection of simulated microcalcifications on mammograms using hardcopy images

Chao Jen Lai, Chris C. Shaw, Gary J. Whitman, Wei T. Yang, Peter J. Dempsey, Victoria Nguyen, Mary F. Ice

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

The aim of this study was to compare mammography systems based on three different detectors - a conventional screen-film (SF) combination, an a-Si/CsI flat-panel (FP)-based detector, and a charge-coupled device (CCD)-based x-ray phosphor-based detector - for their performance in detecting simulated microcalcifications (MCs). 112-150 νm calcium carbonate grains were used to simulate MCs and were overlapped with a slab phantom of simulated 50% adipose/50% glandular breast tissue-equivalent material referred to as the uniform background. For the tissue structure background, 200-250 νm calcium carbonate grains were used and overlapped with an anthropomorphic breast phantom. All MC phantom images were acquired with and without magnification (1.8X). The hardcopy images were reviewed by five mammographers. A five-point confidence level rating was used to score each detection task. Receiver operating characteristic (ROC) analysis was performed, and the areas under the ROC curves (Azs) were used to compare the performances of the three mammography systems under various conditions. The results showed that, with a uniform background and contact images, the FP-based system performed significantly better than the SF and the CCD-based systems. For magnified images with a uniform background, the SF and the FP-based systems performed equally well and significantly better than the CCD-based system. With tissue structure background and contact images, the SF system performed significantly better than the FP and the CCD-based systems. With magnified images and a tissue structure background, the SF and the CCD-based systems performed equally well and significantly better than the FP-based system. In the detection of MCs in the fibroglandular and the heterogeneously dense regions, no significant differences were found except that the SF system performed significantly better than the CCD-based system in the fibroglandular regions for the contact images.

Original languageEnglish (US)
Article number002
Pages (from-to)3901-3919
Number of pages19
JournalPhysics in medicine and biology
Volume51
Issue number16
DOIs
StatePublished - Aug 21 2006

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Receiver operating characteristic analysis for the detection of simulated microcalcifications on mammograms using hardcopy images'. Together they form a unique fingerprint.

Cite this