Abstract
In this study, the Plaque Simulator™ eye plaque brachytherapy planning system was commissioned for ROPES eye plaques and Amersham Health model 6711 Iodine 125 seeds, using TG43-UI data. The brachytherapy module of the RADCALC® independent checking program was configured to allow verification of the accuracy of the dose calculated by Plaque Simulator™. Central axis depth dose distributions were compared and observed to agree to within 2% for all ROPES plaque models and depths of interest. Experimental measurements were performed with a customized PRESAGEm 3-D type dosimeter to validate the calculated depth dose distributions. Preliminary results have shown the effect of the stainless steel plaque backing decreases the measured fluorescence intensity by up to 25%, and 40% for the 15 mm and 10 mm diameter ROPES plaques respectively. This effect, once fully quantified should be accounted for in the Plaque Simulator™ eye plaque brachytherapy planning system.
Original language | English (US) |
---|---|
Article number | 012102 |
Journal | Journal of Physics: Conference Series |
Volume | 444 |
Issue number | 1 |
DOIs | |
State | Published - 2013 |
Event | 7th International Conference on 3D Radiation Dosimetry, IC3DDose 2012 - Sydney, NSW, Australia Duration: Dec 4 2012 → Dec 8 2012 |
ASJC Scopus subject areas
- General Physics and Astronomy