Spot scanning proton therapy minimizes neutron dose in the setting of radiation therapy administered during pregnancy

Research output: Contribution to journalArticlepeer-review

15 Scopus citations

Abstract

This is a real case study to minimize the neutron dose equivalent (H) to a fetus using spot scanning proton beams with favorable beam energies and angles. Minimum neutron dose exposure to the fetus was achieved with iterative planning under the guidance of neutron H measurement. Two highly conformal treatment plans, each with three spot scanning beams, were planned to treat a 25-year-old pregnant female with aggressive recurrent chordoma of the base of skull who elected not to proceed with termination. Each plan was scheduled for delivery every other day for robust target coverage. Neutron H to the fetus was measured using a REM500 neutron survey meter placed at the fetus position of a patient simulating phantom. 4.1 and 44.1 μSv/fraction were measured for the two initial plans. A vertex beam with higher energy and the fetal position closer to its central axis was the cause for the plan that produced an order higher neutron H. Replacing the vertex beam with a lateral beam reduced neutron H to be comparable with the other plan. For a prescription of 70 Gy in 35 fractions, the total neutron H to the fetus was estimated to be 0.35 mSv based on final measurement in single fraction. In comparison, the passive scattering proton plan and photon plan had an estimation of 26 and 70 mSv, respectively, for this case. While radiation therapy in pregnant patients should be avoided if at all possible, our work demonstrated spot scanning beam limited the total neutron H to the fetus an order lower than the suggested 5 mSv regulation threshold. It is far superior than passive scattering beam and careful beam selection with lower energy and keeping fetus further away from beam axis are essential in minimizing the fetus neutron exposure.

Original languageEnglish (US)
Pages (from-to)366-376
Number of pages11
JournalJournal of applied clinical medical physics
Volume17
Issue number5
DOIs
StatePublished - 2016

Keywords

  • Fetus
  • Neutron dose equivalent
  • Proton therapy
  • Spot scanning

ASJC Scopus subject areas

  • Radiation
  • Instrumentation
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Spot scanning proton therapy minimizes neutron dose in the setting of radiation therapy administered during pregnancy'. Together they form a unique fingerprint.

Cite this