Stability of a mobile electron linear accelerator system for intraoperative radiation therapy

Research output: Contribution to journalArticlepeer-review

15 Scopus citations

Abstract

The flexibility of mobile electron accelerators, which are designed to be transported to an operating room and plugged into a normal 3-phase outlet, make them ideal for use in intraoperative radiation therapy. However, their transportability may cause trepidation among potential users, who may question the stability of such an accelerator over a period of use. In order to address this issue, we have studied the short-term stability of the Mobetron system over 20 daily quality assurance trials. Variations in output generally varied within ±2% for the four energies produced by the unit (4, 6, 9, and 12 MeV) and changes in energy produced an equivalent shift of less than 1 mm on the depth-dose curve. Hours of inactivity, with the Mobetron powered on for use either throughout the day or overnight, led to variations in output of about 1%. Finally, we have tested the long-term stability of the absolute dose output of the Mobetron, which showed a change of about 1% per year.

Original languageEnglish (US)
Pages (from-to)3128-3131
Number of pages4
JournalMedical physics
Volume32
Issue number10
DOIs
StatePublished - Oct 2005