Validation of a method for estimating peak skin dose from CT-guided procedures

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

A method for estimating peak skin dose (PSD) from CTDIvol has been published but not validated. The objective of this study was to validate this method during CT-guided ablation procedures. Radiochromic film was calibrated and used to measure PSD. Sixty-eight patients were enrolled in this study, and measured PSD were collected for 46 procedures. CTDIvol stratified by axial and helical scanning was used to calculate an estimate of PSD using the method [1.2 × CTDIvol(helical) + 0.6 × CTDIvol(axial)], and both calculated PSD and total CTDIvol were compared to measured PSD using paired t-tests on the log-transformed data and Bland-Altman analysis. Calculated PSD were significantly different from measured PSD (P < 0.0001, bias, 18.3%, 95% limits of agreement, −63.0% to 26.4%). Measured PSD were not significantly different from total CTDIvol (P = 0.27, bias, 3.97%, 95% limits of agreement, −51.6% to 43.7%). Considering that CTDIvol is reported on the console of all CT scanners, is not stratified by axial and helical scanning modes, and is immediately available to the operator during CT-guided interventional procedures, it may be reasonable to use the scanner-reported CTDIvol as an indicator of PSD during CT-guided procedures. However, further validation is required for other models of CT scanner.

Original languageEnglish (US)
Pages (from-to)224-228
Number of pages5
JournalJournal of applied clinical medical physics
Volume22
Issue number6
DOIs
StatePublished - Jun 2021

Keywords

  • CT dosimetry
  • CT-guided intervention
  • CTDI
  • ablation
  • peak skin dose

ASJC Scopus subject areas

  • Radiation
  • Instrumentation
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Validation of a method for estimating peak skin dose from CT-guided procedures'. Together they form a unique fingerprint.

Cite this